Dissolved Oxygen Controller

Operation

Teledyne Analytical Instruments

47

2DO

8.40 ppm

Cal

-173.8 mV/FS


Span calibration menu pH

3. Clean the electrode and insert the sensor into the desired buffer.
4. Press both horizontal CALIBRATE keys simultaneously to enter

the calibrate mode. Observe the diagnostic value on the bottom
line change from historical data to a real-time value and that the
term "CAL" has disappeared.

5. To change the buffer point, position the cursor under the desired

digit using the left-hand CALIBRATE key.

6. Increase or decrease the value by pressing the appropriate

vertical CALIBRATE key.

7. Wait for the real-time mV value to stabilize.
8. To save the calibration and exit the calibrate mode, press either

MENU SELECTION key.

A perfect electrode slope (efficiency) is typically greater than -150.

mV/FS. If the value falls below -100.0 mV/FS, the electrode should be
serviced or replaced.

3.4.2.3 T

WO

-P

OINT

B

UFFER

C

ALIBRATION

(S

PAN

)

When first installing an electrode, the second cal point

establishes the slope variations or efficiency. The span is typically done
in air which is the saturation point of the solution when in equilibrium.


Procedure:

1. Perform a one-point calibration as outlined in Section 3.4.2.2.
2. Press the appropriate MENU SELECTION key to reach the

Electrode Span Menu and locate the cursor under the "2" as
illustrated.