Dissolved Oxygen Controller

Installation

Teledyne Analytical Instruments

21

2.3.1 AC Power

The LXT-220 requires a 115 VAC 60 Hz power source. A 24VDC

unit is available as an option.

CAUTION: MAKE SURE THE CIRCUIT BREAKER PROTECTING

THE AC INPUT LINE IS OFF BEFORE WIRING THE
LXT-220.

Application drawing 4024003 illustrates the proper power wiring

for single channel LXT-220 as does the affixed label on the backside of
the front panel. The wires are best fed through the right conduit as
shown in Figure 2-2.

Figure 2-2: LXT-220 Electrical Connections

2.3.2 Sensor Wiring

A conditioned input is required from the sensor or electrode to the

LXT-220 for proper operation. Teledyne’s SP1 and SP2 sensors will
provide the required conditioning. The application drawing 4024003