beautypg.com

Teledyne LXT-220 User Manual

Page 14

background image

Introduction

LXT-220 DO

Teledyne Analytical Instruments

14

The LXT-220 can be upgraded with pre-pHault Diagnostics &

HART Communications option which alerts the operator to a pending
DO, pH, ORP, or ISE electrode failure.

The LXT-220 is housed in a rugged NEMA-4X housing with all

operator switches conveniently located on the front panel.

1.2 Main Features of the Transmitter

The LXT-220 Controller for dissolved oxygen applications

incorporates the following standard features:

• Front panel membrane switches

• Menu driven backlit display with contrast adjustment

• Automatic Calibration. This allows calibration points to be

defined upon initialization of the transmitter.

• An easy back to factory calibration menu allows the instrument

to return to a predetermined factory calibration. The default
parameters are: zero electrode offset and –157.5mV/FS slope.

• 4-20mA current output fully expandable to any range within the

transmitter’s full operating range. The output is field selectable
and fully programmable. The current output can be configured to
be reverse acting.

• Temperature Conversion. Temperature can be displayed in either

Fahrenheit or Celsius. The units can be toggled from one to the
other in the field with the conversion calculation performed
automatically.

1.3 Optional Features

The Model LXT-220 DO Controller is a versatile instrument

designed to meet the needs for liquid analytics and process control
across a wide range of applications. There are many options available
for the transmitter including:

Dual Input: An optional second input can be provided for pH,

ORP, resistivity/conductivity, or Specific Ion. The second input
can also be used as a second analysis for dissolved oxygen or as
specific compensation for the primary input.