Installation

LXT-220 DO

Teledyne Analytical Instruments

20

• Handrail mounting plate

• Panel mounting configurations.

U-bolts may be ordered separately. Refer to the following

application drawings for mounting details:

4024011 Dimensions, wall & pipe mounting.
The unit is supplied with a universal-mounting bracket. Two slots
on each side will accommodate U-Bolts for pipe mounting.
4024012 Dimensions, Panel Mounting. This installation uses 1/4"
screws. Panel mounting is accommodated using the captive
jackscrew that will cinch the instrument bezel against the panel.
2000268 Dimensions, handrail mounting.
These application drawings can be located in the back of this

manual.

2.3 Electrical Connections

All electrical connections are made with connectors to a panel

inside the NEMA-4X enclosure. To gain access to the panel, use a flat-
blade screwdriver to remove the 4 captive screws on the front of the
instrument and swing the panel out. See Figure 2-1.

Figure 2-1: Opening the Front Panel