Cyclo hexane C

6

H

12

-1.56

Nitric

oxide

NO

+40.00

Ethane C

2

H

6

-0.43

n-Octane

C

8

H

18

-2.50

Ethylene C

2

H

4

-0.26

n-Pentane

C

5

H

12

-1.45

Helium He

+0.30

Propane C

3

H

8

-0.86

n-Heptane C

7

H

16

-2.10

Propylene

C

3

H

6

-0.55

n-Hexane C

6

H

14

-1.70

Vinyl

chloride

-0.63

Hydrogen H

2

+0.24

Water

H

2

O -0.02

Hydrogen bromide HBr

-0.61

Xenon Xe

-0.95

6.3 Oven and Temperature Controller

The oven in the Analysis Unit is used to provide thermal stability to

the sensor assembly is comprised of dual 40 watt heaters powered by 115
or 230 V ac from the Control Unit. The oven temperature is maintained at
40

°

C by a PID temperature controller housed in the Control Unit. An RTD

sensor is mounted in the Analysis Unit to provide feedback to the PID
controller. The temperature in the Analysis Unit at any time can be read
from the display on the temperature controller after opening the front
panel of the Control Unit. See Figure 6-5. This can be done without
affecting the thermal environment of the Analysis Unit.

Figure 6-5: Access to Temperature Controller in Control Unit