Paramagnetic O2 Purity Analyzer

Operation

Teledyne Analytical Instruments

57

4.7 The Analyze Function

Normally, all of the functions automatically switch back to the

Analyze

function when they have completed their assigned operations.

Pressing the

Escape

button in many cases also switches the analyzer back

to the

Analyze

function. Alternatively, you can press the

Analyze

button

at any time to return to analyzing your sample.

4.8 Signal Output

The standard Model 3010MA Paramagnetic Oxygen Purity Analyzer

is equipped with two 0–1 V dc analog output terminals accessible on the
back panel (one concentration and one range ID). The standard model also
includes one isolated 4–20 mA dc current outputs for the oxygen
concentration.

See Rear Panel in Chapter 3, Installation, for illustration.

The signal output for concentration is linear over the currently

selected analysis range. For example, if the analyzer is set on range that
was defined as 98–100% O

2

, then the output would be:

Table 4.1:Linear Analog Output

% Oxygen

Voltage Signal

Current Signal

Output (V dc)

Output (mA dc)

98.0 0

4

98.2 0.1

5.6

98.4 0.2

7.2

98.6 0.3

8.8

98.8 0.4

10.4

99.0 0.5

12

99.2 0.6

13.6

99.4 0.7

15.2

99.6 0.8

16.8

99.8 0.9

18.4

100 1.0

20