Part 1 Control Unit

Model 3010MA

Teledyne Analytical Instruments

60

2. To change between American and European fuses, remove the

single retaining screw, flip Fuse Block over 180 degrees, and
replace screw.

3. Replace fuse as shown in Figu re 5-2.

4. Reassemble Housing as shown in Figure 5-1.

American Fuses

European Fuses

Figure 5-2: Installing Fuses

5.3 System Self Diagnostic Test

1. Press the

System

button to enter the system mode.

2. Use the < > arrow keys to move to More, and press

Enter.

3. Use the < > arrow keys to move to Self-Test, and press

Enter.

The following failure codes apply:

Table 5-1: Self-Test Failure Codes

Power

0 OK

1 5

V

Failure

2 15

V

Failure

3 Both

failed

Analog

0 OK

1

D/A A (0-1 V Concentration)

2

D/A B (0-1 V Range ID)

3 Both

Failed