Paramagnetic O2 Purity Analyzer

Operation

Teledyne Analytical Instruments

41

Operation

4.1 Introduction

Once the analyzer has been installed, it can be configured for your

application. To do this you will:

1. Set

system

parameters:

2. Establish a security password, if desired, requiring the operator

to log in.

3. Establish and start an automatic calibration cycle, if desired.

4. Calibrate the instrument.

5. Accept the default configuration or define the three user

selectable analysis ranges. Then choose autoranging or select a
fixed range of analysis, as required.

6. Set alarm setpoints, and modes of alarm operation (latching,

failsafe, etc).

Before you configure your 3010MA these default values are in effect:

Ranges: LO = 95-100 %, MED = 96-100 %, HI = 98-100 %
Auto Ranging: ON
Alarm Relays: Defeated, 90%, HI, Not failsafe, Not latching
Zero: Auto, every 0 days at 0 hours, at 95% O

2

Span: Auto, at 100% O2 every 0 days at 0 hours

If you choose not to use password protection, the default password is

automatically displayed on the password screen when you start up, and
you simply press

Enter

for access to all functions of the analyzer.

4.2 Using the Data Entry and Function Buttons

Data Entry Buttons: The < > arrow buttons select options from the menu
currently displayed on the VFD screen. The selected option will blink.

When the selected option includes a modifiable item, the

∆∇

arrow

buttons can be used to increment or decrement that modifiable item.