Paramagnetic O2 Purity Analyzer
Operation
Teledyne Analytical Instruments
41
Operation
4.1 Introduction
Once the analyzer has been installed, it can be configured for your
application. To do this you will:
1. Set
system
parameters:
2. Establish a security password, if desired, requiring the operator
to log in.
3. Establish and start an automatic calibration cycle, if desired.
4. Calibrate the instrument.
5. Accept the default configuration or define the three user
selectable analysis ranges. Then choose autoranging or select a
fixed range of analysis, as required.
6. Set alarm setpoints, and modes of alarm operation (latching,
failsafe, etc).
Before you configure your 3010MA these default values are in effect:
Ranges: LO = 95-100 %, MED = 96-100 %, HI = 98-100 %
Auto Ranging: ON
Alarm Relays: Defeated, 90%, HI, Not failsafe, Not latching
Zero: Auto, every 0 days at 0 hours, at 95% O
2
Span: Auto, at 100% O2 every 0 days at 0 hours
If you choose not to use password protection, the default password is
automatically displayed on the password screen when you start up, and
you simply press
Enter
for access to all functions of the analyzer.
4.2 Using the Data Entry and Function Buttons
Data Entry Buttons: The < > arrow buttons select options from the menu
currently displayed on the VFD screen. The selected option will blink.
When the selected option includes a modifiable item, the
∆∇
arrow
buttons can be used to increment or decrement that modifiable item.