Paramagnetic O2 Purity Analyzer

Operation

Teledyne Analytical Instruments

49

4.3.5 System Self-Diagnostic Test

The Model 3010MA has a built-in self-diagnostic testing routine.

Pre-programmed signals are sent through the power supply, output board
and sensor circuit. The return signal is analyzed, and at the end of the test
the status of each function is displayed on the screen, either as OK or as a
number between 1 and 3. (See System Self Diagnostic Test in Chapter 5
for number code.)

The self diagnostics are run automatically by the analyzer whenever

the instrument is turned on, but the test can also be run by the operator at
will. To initiate a self diagnostic test during operation:

Press the

System

button to start the

System

function.

Contrast Auto—Cal

PSWD Logout More

Use the < > arrow keys to blink More, then press

Enter

.

Version Self—Test

Sensor Output: ###

Use the < > arrow keys again to move the blinking to the Self–Test

function. The screen will follow the running of the diagnostic.

RUNNING DIAGNOSTIC

Testing Preamp — 83

During preamp testing there is a countdown in the lo wer right corner

of the screen. When the testing is complete, the results are displayed.

Power: OK

Analog: OK

Preamp: 0

The module is functioning properly if it is followed by OK. A number

indicates a problem in a specific area of the instrument. Refer to Chapter 5
Maintenance and Troubleshooting
and Table 5-1 for number-code
information. The results screen alternates for a time with:

Press Any Key

To Continue...

Then the analyzer returns to the initial System screen.