Trace Nitrogen in Argon Analyzer

Maintenance

Teledyne Analytical Instruments

67

American Fuses

European Fuses

Figure 5-2: Installing Fuses

5.4 System Self Diagnostic Test

1. Press the System button to enter the system mode.
2. Use the ◄► arrow keys to move to More, and press Enter.
3. Use the ◄► arrow keys to move to Self-Test, and press

Enter.

The following failure codes apply:
Power: Tests the sample 5V and 15V power supplies

0

OK

1

5

V

Failure

2

15

V

Failure

3

Both

Failed

Analog: Set non-zero sample on analog output channel 1 (% of range)
and channel 2 (Range ID) and sample response.

0

OK

1

Analog

channel

1

(%

of Range) is incorrect

2

Analog

channel

1

(Range ID) is incorrect

3

Both

channels

are

incorrect

Preamp: A) Tests the bypass sensor input.
B) Sets a zero level on preamp then tests response.
C)

Sets a non-zero livel on preamp then tests response.

0

OK

1

Preamp is incorrect at zero level

2

Preamp is incorrect at non-zero level

3

Both

zero

and

non-zero are incorrect