beautypg.com

Teledyne 2120XL - Trace Nitrogen Analyzer User Manual

Page 67

background image

Trace Nitrogen in Argon Analyzer

Operation

Teledyne Analytical Instruments

55

regulator on the sample inlet. Do not exceed a maximum of
20 psig (138 kPa)
.

4.5.2.2

R

ECOMMENDED

C

ALIBRATION

G

ASES

Teledyne recommends the following gases be used for calibration:

Zero Gas:

N

2

-free Ar

Span Gas:

For 0-1 ppm range: 0.7-1.0

ppm

N

2

in Ar

For 0-10 ppm range: 7-10 ppm N

2

in Ar

For 0-100 ppm range: 70 – 100 ppm N

2

in Ar (best for overall

accuracy on all ranges)

The first calibration point is fixed by a zero gas. The most

convenient source of nitrogen-free argon is through an activated
molecular sieve scrubber cooled to liquid oxygen temperatures or a
heated transition metal scrubber. A commercially available scrubber the
Teledyne Part number S1748 is ideal for this. Alternatively, a zero gas
cylinder with a low ppb nitrogen level can be used for zeroing in the low
ppm range specified, as long as the N

2

concentration is known.

The second point is fixed by a certified span gas normally supplied

in a high-pressure cylinder.

4.5.2.3

Z

ERO

C

AL

The ZERO button on the front panel is used to enter the zero

calibration function. Zero calibration can be performed in either the
automatic or manual mode. In the automatic mode, an internal
algorithm compares consecutive readings from the sensor to determine
when the output is within the acceptable range for zero. In the manual
mode, the operator determines when the reading is within the acceptable
range for zero. Make sure the zero gas is connected to the instrument.
See Section 3.6

for gas connections.

After pressing the ZERO button, the mode selection screen appears:

Select Zero

Mode : Manual (Auto)

where Manual and Automatic can be toggled back and forth using the