Operation Model

2120XL

Teledyne Analytical Instruments

56

▲/▼ arrow keys. Once the proper mode is displayed, press ENTER.
The following screen appears:

Zero Offset

0.0

pp

b

to Begin

This screen allows the user to add a zero offset if desired. With the

value adjacent to Zero Offset blinking, use the ▲/▼ arrow keys to enter
any offset required. Press ENTER when finished.

4.5.2.3.1 Manual Zero

If manual zero mode was selected, the following screen appears:

N2: 1.2

pp

b Zero

to Finish

In the Manual Mode, the user has the opportunity to decide when

the zero settling is close enough to zero to be valid. When the value
adjacent to

N2

is sufficiently close to zero and sufficiently stable, press

ENTER to accept the zero reading.

After pressing ENTER two screens will appear that indicate that the

zero level has been acquired and will report the live adc value for the
two gains in the amplifier (G0 and G1) as well as the change (Δ).

Zero Acquired

G1/1

Δ 1.2

Val:-12345

shortly followed by:

Zero Acquired

G0/1

Δ-0.2 Val:-12345

4.5.2.3.2 Auto Mode Zero

If AUTO mode was selected from the Zero Mode Selection screen

the following screen appears:

Zero Settling:

0:58

The time value displayed at the right count down to zero from the

value currently held in Calholdtimer. See Section 4.5.1.9.