Teledyne 2120XL - Trace Nitrogen Analyzer User Manual

Page 64

background image

Operation Model

2120XL

Teledyne Analytical Instruments

52

Repeat the process for adjusting the 20 mA output. Pressing

ENTER will accept the value and return the display to the last System
screen.

4.5.1.12

L

INEARIZE

(L

IN

)

With this function, the user has the capability of linearizing the

output of the instrument by flowing prepared samples with known
concentration across the analytical range. The linearization can be
performed manually or automatically depending on the setting chosen in
the linearization setup screen. Upon completion of the linearization
sample introduction, the software performs linear approximation
between data points.

The Model 2120XL can accept up to 11 linearization points. These

are:

 Zero point (pure argon)
 9 data points between zero and full span
 Full Span (85-100 ppm N

2

in Ar or 85-100% of the highest

full range)

The linearization can be performed manually or automatically

depending on the chosen setting.

To enter the linearization function:
Press the SYSTEM key to display the first System menu. Use the

◄►arrow keys to position the blinking over the More function, and
press ENTER. Repeat the above to ENTER at More a second time. This
will display a third screen of system functions as shown below.

Model Out-cal Lin

Use the ◄►arrow keys to position the blinking over the Lin func-

tion, and press ENTER. The following screen will appear:

Linearize

Manual Auto

4.5.1.12.2 Setup Sub Function

The Setup sub function allows the operator to choose between

Manual or Automatic linearization. In manual mode the operator the
operator enters the value of the prepared linearization sample as reported