Teledyne 2120XL - Trace Nitrogen Analyzer User Manual
Page 36
Installation Model
2120XL
Teledyne Analytical Instruments
24
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that
request.
Once a zero or span command is sent, and acknowledged (contact
closes), release it. If the command is continued until after the zero or
span is complete, the calibration will repeat and the Cal Relay Contact
(CRC) will close again.
For example:
1. Test the CRC. When the CRC is open, Send a zero command
until the CRC closes (The CRC will quickly close.)
2. When the CRC closes, remove the zero command.
3. When CRC opens again, send a span command until the
CRC closes. (The CRC will quickly close.)
4. When the CRC closes, remove the span command.
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The Remote Valve connections (described below) provides
signals to ensure that the zero and span gas valves will be
controlled synchronously. If you have the Auto Caibration
Option, it includes additional zero and span gas inputs—
the 2120XL automatically regulates the zero, span and
sample gas flow.
Range ID Relays: Four dedicated Range ID relay contacts. The first
three ranges are assigned to relays in ascending order—Low range is
assigned to Range 1 ID, Medium range is assigned to Range 2 ID, and
High range is assigned to Range 3 ID. The fourth range is reserved for
the Cal Range. Table 3-4 lists the pin connections.
Table 3-4: Range ID Relay Connections
Pin Function
21
Range 1 ID Contact
38
Range 1 ID Contact
22
Range 2 ID Contact
39
Range 2 ID Contact
19
Range 3 ID Contact