Trace Nitrogen in Argon Analyzer

Operation

Teledyne Analytical Instruments

53

by the analyzer’s display. In Auto mode, this is done automatically by
the software.

From the Linearizer (Lin) function screen, use the ◄►arrow keys

to position the blinking over the Setup function, and press ENTER. The
following screen appears:

Linearize

Manual Auto

To enter Manual or Auto Mode, use the ◄►arrow keys to position

the blinking over the Manual or Auto and press ENTER. The following
screen appears:

In 0.0

pp

b

1

Out 0.0

pp

b Accept

where the number (1) in the right corner of the screen indicates the
linearization sample number currently being entered. In this exampe, it
is the first sample to be entered.

In MANUAL mode, the number adjacent to “

In”

is editable. It

represents the unlinearized concentration of the sample gas as reported
by the analyzer. In AUTO mode, this value is not editable but is entered
automatically by the routine. The value for “

Out

” is editable in both Man

and Auto modes.

Out

is entered by the user and represents the desired

linearized value of the sample. For instance, feeding the analyzer a
prepared linearizer sample gas whose concentration is known to be 10.0
ppm and the analyzer reports an unlinearized value of

13.6pm

adjacent to

In

, the user would correct this by entering

10.0pm

as the

Out

value.

If not done already, hook up the first linearization sample to the

sample input of the analyzer and allow the gas to flow for a few minutes.
Watch the display and when the concentration value is stable, enter the
correct value for

Out

and then press ENTER to accept the value.

A confirmation screen will appear.

Accept

Reject

Pressing ENTER again will save the data into memory to be used

for calculating the coefficients used in the best fit linearization