Dig Filt:

The Model 2120XL includes a user adjustable

digital filter. The filter has settings 1-10. Setting 1 is the least
amount of filtering and 10 is the highest level of damping.

Self–Test

: The instrument performs a self-diagnostic test to

check the integrity of the power supply, output boards and
amplifiers.

PSWD:

Security can be established by choosing a 5 digit

password (PSWD) from the standard ASCII character set.
(See Installing or Changing a Password, below, for a table of
ASCII characters available.)

Logout:

Logging out prevents an unauthorized tampering

with analyzer settings.

More

: Select and enter More to get a new screen with

additional sub functions listed.

Auto-Cal

: Used to define an automatic calibration sequence

and/or start an Auto-Cal.

Flow:

Used to set the flow through the flow controller.

TRAK/HLD

: The operator sets whether the instrument

analog outputs track the concentration change during
calibration and sets a time delay for the concentration alarms
after calibration

Cal Hold Timer:

Allows alarm suppression and analog

output value hold times during calibration events for a user
adjustable period.

Model

: Displays Manufacturer, Model, and Software

Version of the instrument.

Out-Cal:

Analog output calibration allows the user to

calibrate the 4 and 20 mA analog output values.

Lin:

Used to linearize the output of the controller.

To enter any of the system functions, use the ◄►arrow keys to

move to the function and then use the ENTER key to engage that
function.

4.5.1.1

D

IGITAL

F

ILTER

Upon entering the digital filter function, the following screen

appears on the display: