Trace Nitrogen in Argon Analyzer

Installation

Teledyne Analytical Instruments

33

3.8 Calibration

The Model 2120XL was calibrated at the factory in accordance

with the calibration data shipped with your instrument. Prior to using the
instrument for analysis the calibration must be checked using the factory
settings supplied. If the settings do not match, the instrument should be
recalibrated prior to using for the first time.

Calibration procedures require an understanding of how to operate

the instrument which is discussed in Section 4 of this manual. Refer to
Section 4.5.2 for calibration procedures after reading the entire section
on operating the instrument (Section 4).

3.9 Installation Checklist

After installing the system, and prior to powering up the system,

verify the following items:

1. Each gas line has been purged for at least 1 hour prior to

connection to the analyzer inlets.

2. There are no leaks at the sample inlet gas connection to the

instrument.

3. There are no leaks between the span gas cylinder and the

analyzer span gas inlet port when span gas is flowing to the
analyzer.

4. There are no leaks between the zero gas source and the

analyzer zero gas inlet port when zero gas is flowing to the
analyzer.

5. The vent tube connection has no leaks, has an

unobstructed path to atmosphere, and is at least 1/8-in
(0.32-cm) OD.

6. All electrical connections have been made properly and the

Power Entry module is set for the appropriate power from
your facility.