beautypg.com

Teledyne 2120XL - Trace Nitrogen Analyzer User Manual

Page 54

background image

Operation Model

2120XL

Teledyne Analytical Instruments

42

indicates a problem in a specific area of the instrument. Refer to Chapter
5 Maintenance and Troubleshooting for number-code information. The
results screen alternates for a time with:

Press Any Key

To Continue...


Then the analyzer returns to the initial System screen.

4.5.1.3

P

ASSWORD

P

ROTECTION

Security can be established by choosing a 3 digit password (PSWD)

from the standard ASCII character set. (See Installing or Changing a
Password, below, for a table of ASCII characters available.)

Once a unique password is assigned and activated, the operator

MUST enter the UNIQUE password to gain access to set-up functions
which alter the instrument's operation, such as setting the instrument
span or zero setting, adjusting the alarm setpoints, or defining an
analysis range.

After a password is assigned, the operator must log out to activate

it. Until then, anyone can continue to operate the instrument without
entering the new password.

Only one password can be defined. Before a unique password is

assigned, the system assigns TAI by default. This allows access to
anyone. After a unique password is assigned, to defeat the security, the
password must be changed back to TAI.

If a password is assigned, then setting the following instrument

parameters can be done only after the password is entered: span and
zero settings, alarm setpoints, analysis range definitions, switching
between auto ranging and manual override. In addition, all System
functions such as setting up an auto-cal, self-test and assigning a new
password etc. with the exception of Track/Hold and Model cannot be
accessed without entering a password. However, the instrument can still
be used for analysis without entering the password.

If you have decided not to employ password security, use the

default password TAI. This password will be displayed automatically by