Teledyne 2120XL - Trace Nitrogen Analyzer User Manual
Page 30

Installation Model
2120XL
Teledyne Analytical Instruments
18
3.2 Choosing a Location
Locate the Model 2120XL in a clean area free of:
Excessive dust
Mechanical vibrations
Strong electric or electromagnetic fields
Corrosive gases
Moisture exceeding 90% relative humidity
The use of walkie-talkies or cellular phones
Choose a location where sudden temperature changes in excess of
5°F (5°C) do not occur and where the temperature does not exceed the
specified ambient temperature range. Avoid any location where the
instrument would be exposed to direct sunlight or radiation from heaters.
CAUTION:
THIS ANALYZER IS DESIGNED FOR USE IN A
GENERAL PURPOSE AREA AND IS NOT RATED
FOR USE IN HAZARDOUS AREAS.
3.3 Mounting
The Model 2120XL is a rack mountable unit intended for indoor
use only. Space and materials required for mounting are:
8-3/4-inches (22.2 cm) of 19-inch rack space.
Service access space behind and in front of the analyzer.
Four mounting screws suited to the rack.
All operator controls are mounted on the control panel, which is
hinged on the left edge and doubles as the door that provides access to
components inside the controller module. The door latch is spring
loaded and will swing open when the button in the center of the latch
(upper right corner) is pressed all the way in with a narrow gauge tool
(less than 0.18 inch wide). Allow clearance for the door to open in a
90-degree arc of radius 7.125 inches. See Figure 3-1.