Operation Model

2120XL

Teledyne Analytical Instruments

46

4.5.1.4

L

OGOUT

The Logout function provides a convenient means of leaving the

analyzer in a password protected mode without having to shut the
instrument off. By entering Logout, you effectively log off the
instrument leaving the system protected against use until the password is
reentered. To log out, press the System button to enter the System
function.

Dig_filt Self-test

Pwd Logout Err More

Use the ◄►arrow keys to position the blinking over the Logout

function, and press ENTER to Log out. The screen will display the
message:

Protected Until

Password Reentered

4.5.1.5

ERR

During analysis, if the system detects a fault, and error signal is

sent and shows a flashing ‘E’ in the right corner of the Analyze screen
as follows:

N2 2.3

pp

b E

Rng: 100.0pb 100.2

cc

m

To obtain more information about the error, press SYSTEM and

navigate to the Err function. Then pres ENTER to display the particular
error screen. There are 14 possible error screens that may appear. The
first line displays the particular fault while the second line informs the
user to press ENTER to clear the error.

Note: The particular fault must be remedied before attempting to

clear the error, otherwise the error indication will
immediately regenerate on the Analyze screen.

The following notifications may appear on the Err screen first line:

 LED Display Error
 Flow Error