Trace Nitrogen in Argon Analyzer

Operation

Teledyne Analytical Instruments

63

70 0.7 15.2
80 0.8 16.8
90 0.9 18.4

100 1.0 20.0

The analog output signal has a voltage which depends on the

nitrogen concentration AND the currently activated analysis range. To
relate the signal output to the actual concentration, it is necessary to
know what range the instrument is currently on, especially when the
analyzer is in the auto-ranging mode.

To provide an indication of the range, a second pair of analog

output terminals are used. They generate a steady preset voltage (or
current when using the current outputs) to represent a particular range.
The following table gives the range ID output for each analysis range:

Range

Voltage (V)

Current (mA)

LO 0.25

8

MED 0.50

12

HI 0.75 16