Operation Model

2120XL

Teledyne Analytical Instruments

58

▲/▼ arrow keys. Once the proper mode is displayed, press ENTER.
The following screen appears:

Span

100.0

pp

m

to begin Span

This screen allows the user to input the known concentration of the

span gas used. Uuse the ▲/▼ arrow keys to enter the span gas
concentration. Press ENTER when finished.

4.5.2.3.1 Manual Span

If manual span mode was selected, the following screen appears:

39.25

pp

m

to Finish

In the Manual Mode, the user has the opportunity to decide when

the span settling is close enough to the actual concentration to be valid.
When the value is sufficiently close to the known concentration and
sufficiently stable, press ENTER to accept the span calibration.

4.5.2.3.2 Auto Mode Span

If AUTO mode was selected from the Span Mode Selection screen

the following screen appears:

Span Settling:

0:58

The time value displayed at the right count down to zero from the

value currently held in Calholdtimer. See Section 4.5.1.9.

When the counter reaches zero, the display switches to display the

live adc change count.

N2: 123.0

pp

m

Δ 123

Span

where Δ 123.0 is a live display of the change in the adc.

When the change in adc value is reasonably small, the span

calibration is complete and the screen reverts to the Analyze mode
screen.