Thermal Conductivity Analyzer

Part II: Analysis Unit

Teledyne Analytical Instruments

Part II: 4-3

3. Replace fuse by reversing process in step 1.

4.4

System Self Diagnostic Test

1. Press the

System

button to enter the system mode.

2. Use the < > arrow keys to move to

More, and press

Enter

.

3. Use the < > arrow keys to move to

Self-Test, and press

Enter

.

4. Observe the error-code readings on the VFD Display screen, and

check Table 4-1, below, to interpret the codes.

Table 4-1: Self Test Failure Codes

Power

0

OK

1

5 V Failure

2

15 V Failure

3

Both Failed

Analog

0

OK

1

DAC A (0–1 V Concentration)

2

DAC B (0–1 V Range ID)

3

Both Failed

Preamp

0

OK - > 0 means gains of the amplifier have a

high offset, e.g. 204 means the highest
gain has a high offset it is common on
sealed air reference.