beautypg.com

Teledyne 2010B - Split architecture thermal conductivity analyzer User Manual

Page 29

background image

Teledyne Analytical Instruments

Thermal Conductivity Analyzer

Part I: Control Unit

Part I 3-7

Actuates when self test fails.

(Reset by pressing

I/O

button to remove power. Then

press

I/O

again and any other button EXCEPT

System

to resume.

Further detail can be found in chapter 4, section 4-5.

Figure 3-5: Types of Relay Contacts

3.3.5 Digital Remote Cal Inputs

Accept 0 V (off) or 24 V dc (on) inputs for remote control of calibra-

tion. (See Remote Calibration Protocol below.)

Zero:

Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the

Zero

mode. Either side may be

grounded at the source of the signal. A synchronous signal
must open and close the external gas control valves appropri-
ately. See 3.3.9 Remote Probe Connector. (With the –C
option, the internal valves operate automatically.)

Span:

Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the

Span

mode. Either side may be

grounded at the source of the signal. A synchronous signal
must open and close the external gas control valves appropri-
ately. See 3.3.9 Remote Probe Connector. (With the –C
option, the internal valves operate automatically.)

Cal Contact: This relay contact is closed while analyzer is spanning

and/or zeroing. (See Remote Calibration Protocol below.)

Remote Calibration Protocol: To properly time the Digital Remote

Cal Inputs to the Model 2010B Analyzer, the customer's controller must
monitor the Cal Relay Contact.