beautypg.com

Range – Teledyne 2010B - Split architecture thermal conductivity analyzer User Manual

Page 52

background image

4 Operation

Model 2010B

4-18 Part I

Teledyne Analytical Instruments

Use the

∆∆∆∆∆∇

keys to choose between % or ppm units. Then press

Enter

to

move to the next screen.

AL1: 1ØØØ ppm HI

Dft:N Fs:N Ltch:N

Five parameters can be changed on this screen:

Value of the alarm setpoint, AL1: ####

Out-of-range direction, HI or LO

Defeated? Dft:Y/N (Yes/No)

Failsafe? Fs:Y/N (Yes/No)

Latching? Ltch:Y/N (Yes/No).

To define the setpoint, use the < > arrow keys to move the
blinking over to AL1: ####. Then use the

∆∇

arrow keys to

change the number. Holding down the key speeds up the
incrementing or decrementing.

To set the other parameters use the < > arrow keys to move the
blinking over to the desired parameter. Then use the

∆∇

arrow

keys to change the parameter.

Once the parameters for alarm 1 have been set, press

Alarms

again, and repeat this procedure for alarm 2 (AL2).

To reset a latched alarm, go to

Dft– and then press either

two

times or

two times. (Toggle it to

Y and then back to N.)

–OR –

Go to

Ltch– and then press either

two times or

two times.

(Toggle it to

N and back to Y.)

4.6

The

Range

Select Function

The

Range

function allows you to manually select the concentration range

of analysis (

MANUAL), or to select automatic range switching (AUTO).

In the

MANUAL screen, you are further allowed to define the high and low

(concentration) limits of each Range, and select a single, fixed range to run.

CAUTION: If this is a linearized application, the new range must

be within the limits previously programmed using the
System function, if linearization is to apply through-
out the range. Furthermore, if the limits are too small
a part (approx 10 % or less) of the originally linear-
ized range, the linearization will be compromised.