4 Operation

Model 2010B

4-22 Part I

Teledyne Analytical Instruments

Escape

button in many cases also switches the analyzer back to the

Analyze

function. Alternatively, you can press the

Analyze

button at any time to return

to analyzing your sample.

The

Analyze

function screen shows the impurity concentration and the

application gases in the first line, and the range in the second line. In the lower
right corner, the abbreviation

Anlz indicates that the analyzer is in the

Analyze

mode. If there is an * before the

Anlz, it indicates that the range is linearized.

1.95 % H2 — N2

nR1:Ø —10 *Anlz

n indicates non inverting range
i indicates inverting range

If the concentration detected is overrange, the first line of the display blinks

continuously.

4.8

Programming

CAUTION: The programming functions of the Set Range and

Curve Algorithm screens are configured at the facto-
ry to the users application specification. These func-
tions should only be reprogrammed by trained,
qualified personnel.

To program, you must:

1. Enter the password, if you are using the analyzer’s password

protection capability.

2. Connect a computer or computer terminal capable of sending an

RS-232 signal to the analyzer RS-232 connector. (See chapter 3
Installation for details). Send the rp command to the analyzer.
OR

For software 1.1.4 or later, turn the instrument off and back on.
While on the introduction screen hold the Analyze key for at least
fifteen seconds. Press the Enter key twice to return to the
Analyze mode.

3. Press the

System

button to start the

System

function.

CONTRAST AUTO—CAL

PWD LOGOUT MORE

Use the < > arrow keys to blink

MORE, then press

Enter

.

Contrast Function is

DISABLED

(Refer to Section 1.6)