4 Operation

Model 2010B

4-6 Part I

Teledyne Analytical Instruments

Use < > arrows to blink

AUTO—CAL

, and press

Enter

. A new screen for

ZERO/SPAN set appears.

ZERO in Ød Øh off

SPAN in Ød Øh off

Press < > arrows to blink

ZERO

(or

SPAN

), then press

Enter

again.

(You won’t be able to set

OFF to ON if a zero interval is entered.) A Span

Every ... (or Zero Every ...) screen appears.

Zero schedule: OFF

Day: Ød Hour: Øh

Use

∆∆∆∆∆∇

arrows to set an interval value, then use < > arrows to move to the

start-time value. Use

∆∆∆∆∆∇

arrows to set a start-time value.

To turn

ON the SPAN and/or ZERO cycles (to activate AUTO–CAL):

Press

System

again, choose

AUTO–CAL

,

and press

Enter

again. When the

ZERO/SPAN values screen appears, use the < > arrows to blink the ZERO
(or

SPAN) and press

Enter

to go to the next screen. Use < > to select

OFF/

ON field. Use

∆∆∆∆∆∇

arrows to set the

OFF/ON field to ON. You can now turn

these fields

ON because there is a nonzero span interval defined.

4.3.3 Password Protection

Before a unique password is assigned, the system assigns

TAI by default.

This password will be displayed automatically. The operator just presses the
Enter key to be allowed total access to the instrument’s features.

If a password is assigned, then setting the following system parameters can

be done only after the password is entered: alarm setpoints, assigning a new
password, range/application selections, and curve algorithm linearization.
(APPLICATION and ALGORITHM are covered in the programming section.)
However, the instrument can still be used for analysis or for initiating a self-test
without entering the password. To defeat security the password must be
changed back to

TAI.

NOTE: If you use password security, it is advisable to keep a copy of

the password in a separate, safe location.