Operation – Teledyne 2010B - Split architecture thermal conductivity analyzer User Manual

Page 83

background image

Thermal Conductivity Analyzer

Part II: Analysis Unit

Teledyne Analytical Instruments

Part II: 3-1

Operation

3.1

System Self Diagnostic Test

The self diagnostics are run automatically by the analyzer whenever the

instrument is turned on, but the test can also be run by the operator at will.
During the test, internal signals are sent through the power supply, output
board and sensor circuit automatically. The return signal is analyzed, and at
the end of the test the status of each function is displayed on the screen,
either as

OK or as a number between 1 and 3. (See Table 4-1 for number

code.)

Note: Remote Probe connector must be connected to the Analysis

Unit, or sensor circuit will not be properly checked.

Instructions for running self diagnostics are repeated here for your

convenience:

1. Press the

System

button to enter the system mode.

2. Use the < > arrow keys to move to

More, and press

Enter

.

3. Use the < > arrow keys to move to

Self-Test, and press

Enter

.

During preamp testing there is a countdown in the lower right corner of

the screen. When the testing is complete, the results are displayed.

Power: OK Analog: OK

Preamp: 3

The following failure codes apply:

Table 4-1: Self Test Failure Codes

Power

0

OK

1

5 V Failure