Teledyne Analytical Instruments

Part II: 1-3

Thermal Conductivity Analyzer

Part II: Analysis Unit

Figure 1-2: Electrical Connector/Control Panel

Power In

Power input terminals for electric heater.
Requires 110 or 220 V ac, depending on
position of the Voltage Selector switch. Use
50/60 Hz.

CAUTION: Check the position of the Voltage Selector switch

BEFORE applying power to the Power Input termi-
nals.

Voltage Selector Power input selector switch for electric heater.

Adjusts input requirement for 110 or
220 V ac, depending on available source
voltage. Use 50/60 Hz.

Fuses

1.6 A, 250 V, T type, European size
5

×

20 mm fuses. Fuse 2 is on the neutral side

of the line. Fuse 1 is on the hot side of the
line.

Solenoid Valves Terminals that provide all electrical intercon-

nections from the Control Unit to the gas
control valves.

Sensor Signal

Terminals that provide connections from the
T/C sensor to the Control Unit.