Teledyne 2010B - Split architecture thermal conductivity analyzer User Manual

Page 27

background image

Teledyne Analytical Instruments

Thermal Conductivity Analyzer

Part I: Control Unit

Part I 3-5

Figure 3-4: Analog Output Connections

Examples:

The analog output signal has a voltage which depends on gas concen-

tration relative to the full scale of the range. To relate the signal output to the
actual concentration, it is necessary to know what range the instrument is
currently on, especially when the analyzer is in the autoranging mode.

The signal output for concentration is linear over the currently selected

analysis range. For example, if the analyzer is set on a range that was defined
as 0–10 % hydrogen, then the output would be as shown in Table 3-1.

Table 3-1: Analog Concentration Output—Example

Percent

Voltage Signal

Current Signal

Hydrogen

Output (V dc)

Output (mA dc)

0

0.0

4.0

1

0.1

5.6

2

0.2

7.2

3

0.3

8.8

4

0.4

10.4

5

0.5

12.0

6

0.6

13.6

7

0.7

15.2

8

0.8

16.8

9

0.9

18.4

10

1.0

20.0

(Continued)