beautypg.com

Teledyne 2010B - Split architecture thermal conductivity analyzer User Manual

Page 46

background image

4 Operation

Model 2010B

4-12 Part I

Teledyne Analytical Instruments

The analyzer is calibrated using reference, zero, and span gases. Gas

requirements are covered in detail in chapter 3, section 3.4 Gas Connections.
Check that calibration gases are connected to the analyzer according to the
instructions in section 3.4, observing all the prescribed precautions.

Note: Shut off the gas pressure before connecting it to the analyzer,

and be sure to limit pressure to 40 psig or less when turning it
back on.

Readjust the gas pressure into the analyzer until the flowrate through the

sensor settles between 50 to 200 cc/min (approximately 0.1 to 0.4 scfh).

Note: Always keep the zero calibration gases flow as close to the

flowrate of sample gas as possible

4.4.1 Zero Cal

The

Zero

button on the front panel is used to enter the zero calibration

function. Zero calibration can be performed in either the automatic or manual
mode.

CAUTION: If you are zeroing the Cal Range by itself (multiple

application analyzers only), use

manual mode

zeroing.

If you want to calibrate ALL of the ranges at once
(multiple application analyzers only), use

auto mode

zeroing in the Cal Range.

Make sure the zero gas is flowing to the instrument. If you get a

CELL

CANNOT BE BALANCED message while zeroing skip to section 4.4.1.3.

4.4.1.1

Auto Mode Zeroing

Observe the precautions in sections 4.4 and 4.4.1, above. Press

Zero

to

enter the zero function mode. The screen allows you to select whether the zero
calibration is to be performed automatically or manually. Use the

∆∇

arrow

keys to toggle between

AUTO and MAN zero settling. Stop when AUTO

appears, blinking, on the display.

Select zero

mode: AUTO

Press

Enter

to begin zeroing.

####.## % H2 — N2

Slope=#.### C—Zero