beautypg.com

System design considerations, Design for calibration – Liquid Controls MS Meters User Manual

Page 8

background image

8

Mobile Master Proving Trolley

Proper Meter Installation

Volumetric Prover

System Design Considerations

Design For Calibration

Field calibration of meters is essential. Calibration
conducted during meter production is to confirm the meter
is capable of attaining the required accuracy/linearity and
repeatability.

Typically, the meter characteristics are:

Mechanical Registration

Repeatability: Capable of .02% or better at any flow
rate over entire range

Linearity: ± .125% over a 5:1 range

Linearity: ± .22% over a 10:1 range

Linearity: ± .5% over a 40:1 range

Electronic Registration

Repeatability: Capable of .02% or better at any flow
rate over entire range

Linearity: ± .1% over a 5:1 range

Linearity: ± .1% over a 10:1 range

Linearity: ± .15% over a 40:1 range

The meters are normally tested against a master meter
that has been proven against a Weights & Measures
certified, volumetric prover.

The pictures below show a properly configured meter
installation and mobile master prover. The prover
connections are at the right. The two valves to the left of
the meter are for connecting the volumetric prover or

master meter to be used for calibration. The valve
mounted in the vertical line between the two horizontally
mounted valves is a double-block, bleed valve used to
ensure there is no leakage through the valve during
calibration.