Input Type

Jumper Position

E2

E4

E6

E5

Single-ended (signal & return)

a, c

a, c

a, c

c

Complementary (differential)

b

b

b

c

Complementary (with excitation

b

b

-

b, d

and no zero index)

Input Termination

Jumper Position

(for differential inputs only)

E1

E3

E5

For long input cable runs

a

a a

For short input cable runs None

None None

Note: Do not install E5-a if E5-b and d are installed

Input Phase

Jumper Position

E7

Up count, negative transition of B,
A positive ( A leads B )

None

Up count, positive transition of B,
A positive ( B leads A )

a

Count by options

Jumper Position

E9

X1 = positive edge of A input

None

X2 = positive and negative edges of A input

a

X4 = positive & negative edges of both A & B inputs

b

13.

QUADRATURE SIGNAL CONDITIONER

13.1

SETTING JUMPERS FOR INPUT SIGNAL LEVELS

See "ZERO INDEX SETUP" for jumper locations for zero index signal (Page 35).

-31-

1

12

a

c

E1

E2

E3

E5

E7

E4

E6

E8

E9

E10

c

c

c

a

b

c

a

a

a

a

a

a

a

b

b

b

b

b

a

b

c

d