MENU KEY
DIGIT SELECT KEY
VALUE SELECT KEY
Press the
key until SEr
2 (Serial interface setup 2) is
displayed.
Press to display status.
Press again to select digit.
Digit 2: Line feed
Digit 3: Alarm data sent with
meter readings
Digit 4: Control of output
Digit 5: Meter address
Press to select value for
flashing digit
2: "0"-no line feed
"1"-
3: "0"-no alarm data
"1"-alarm data sent
4: "0"-continuous output
"1"-output on command
5: "1" to "F" = 1 to 15
"0." to "F." = 16 to 31
Meter #16 to Meter #31
1 2 3 4 5 6
1 2 3 4 5 6
Press the
key until SEr
3 (Serial interface setup 3) is
displayed.
Press to display status.
Press again to select digit.
Digit 2: RS485 full or half
duplex
Digit 3: Meter recognition
character
Digit 4: RS232 RTS type
Digit 5: Carriage return (and
LF if selected
Digit 6: Data sent via commu-
nications (only 1 item allowed
if BCD)
Press to select value for
flashing digit
2: "0"-Full duplex
"1"-Half duplex
3: "0"-"*" character
"1"-Custom character
4: "0"-Nonlatching RTS
"1"-Latching RTS
5: "0" - End of all items
"1" - At end of each item
6: "0" - All active items sent
"1" - Item 1 only
"2" - Item 2 only (if active)
"3" - Item 3 only (if active)
"4" - Peak value
"5" - Active items +Peak
-43-
1 2 3 4 5
1 2 3 4 5
Press to select value for
flashing digit
3: "0"- Send unfiltered signal
"1"- Send filtered signal
4: "0" to "6" - 300 to 19.2k
5: "0" to "8" 60Hz to rate/256
Press to display status.
Press again to select digit.
Digit 3: Output filtering
Digit 4: Baud rate
Digit 5: Output update rate
Press the
key until SEr 1
(Serial interface setup 1) is
displayed.
15. 2 OPERATING MODE MENU SELECTION
The following menu items are accessible only with an RS-232 or RS-485 option installed and
appropriate lockouts enabled. See Section 9 for further information.