background image

96

Services Menu

13.4 Screen

13.5 Buzzer

13.6 Boards info

13.7 Batteries

13.8 Software download

13.9 End application

Screen

Screen calibration

This screen allows you to adjust the click position corresponding to the
displayed buttons. Use it if ever the sensitive areas are no longer properly
aligned with the buttons on display.
Place the stylus right on the cross when asked and proceed. User will be
asked to press on the screen to accept the new setting. If not, the new
setting is not recorded and no change is made.

Screen brightness

Press the up or down arrow to increase or decrease the screen brightness.
Four levels are pre-set: Max, Comfortable, Standard (default setting) and
Economy

. Press OK when finished.

Note:

This can also be called through the brightness icon on the

main display.

Buzzer

Adjust the sounds available on the instrument.
When “screen tap” is enabled, a click sound is heard each time the screen is touched.
The instrument alarm sound can be enabled or disabled to suit the application. The sound type can also
be adjusted. Press the Test button to test the adjustment made. Press again to stop.

Boards info

Main board info

For reference, this display gives information on the instrument model,
software version and instrument ID.

Measurement board info

For reference, this display gives information on the sensor measurement
board hardware and software.
Pressing the Analog outputs corr. button displays the correction factor that
is applied to the analog outputs.
Pressing the Amplifiers corr. button displays the value of the actual
correction factor on the amplifiers.

Sensor parameters

For reference this display gives information on the sensor model and type,
last calibration, settings and behavior.

Batteries

On all instruments this display gives the real time clock battery charge level and voltage.
Note:

A warning message (and icon) is displayed if the battery level becomes too low and needs to be

replaced.

Software download

For Hach Lange technician use only. Used when reloading the software for new versions.
This ends the application. User must stop and restart the instrument to restart the program.

End application

This ends the application. User must stop and restart the instrument to restart the program.