3 measurement board, 8 measurement alarm relays, Installation 3.7.3 measurement board – Hach-Lange ORBISPHERE K1100_M1100 Operator Manual User Manual

Page 36

background image

34

Installation

3.7.3 Measurement board

3.8 Measurement alarm relays

Figure 17 Measurement board

Figure 18 Connector J7

Connector J7 (inputs & outputs)

Measurement alarms relays:

1.

Common

2.

Output relay 1

3.

Output relay 2

4.

Output relay 3

Analog current outputs:

5.

GND

6.

Output 1

7.

Output 2

8.

Output 3

Digital inputs:

9. to 16.

Not used

Figure 19 Output relays

The three output relays are located on the measurement
board.

They can be individually configured to Normally Open (NO) or
to Normally Closed (NC) by physically moving the jumper on
each relay. On the picture example on the left:

Upper relay is set to NC

Middle relay is set to NO

Lower relay is shown with no jumper

Note:

J14 (upper relay) is relay 1

J18 (middle relay) is relay 2
J19 (lower relay) is relay 3