14
Specifications
Sensor
Sample temperature
Measurement from -5 to 50°C (23 to 122°F)
Sensor resistant to temperature from -5 to 100°C (23 to 212°F)
Sample pressure
1 to 20 bar absolute (14.5 to 290 psia)
Sample flow rate requirement
50 to 300 mL/min
Sample types
K1100 sensor
Water only
M1100 sensor (low range)
Water and beer
M1100 sensor (high range)
Water, beer, wine, wort and carbonated drinks
CIP
K1100 sensor
Not designed for use in CIP processes
M1100 sensor
Resistant to all common CIP methods
Measurement range
Low range sensors: 0 to 2,000 ppb (dissolved)
High range sensors: 0 to 40 ppm (dissolved)
Repeatability
Low range sensors: ± 0.4 ppb or 1%, whichever is the greater
High range sensors: ± 0.015 ppm or ± 2% whichever is the greater
Reproducibility
Low range sensors: ± 0.8 ppb or 2%, whichever is the greater
High range sensors: ± 0.02 ppm or ± 3% whichever is the greater
Accuracy
Low range sensors: ± 0.8 ppb or 2%, whichever is the greater
High range sensors: ± 0.02 ppm or ± 3% whichever is the greater
Limit of detection (LOD)
Low range sensors: 0.6 ppb
High range sensors: 0.015 ppm
Response time (90%)
Low range sensors: < 10 seconds in gas phase; < 30 seconds in liquid phase
High range sensors: < 10 seconds in gas phase; < 50 seconds in liquid phase
Warm-up time
None
Display resolution
0.1 ppb
Calibration
Low range sensors: Single point calibration (zero)
High range sensors: Two at cap replacement (zero and air), one during use (air)
Calibration sample
Low range sensors: Standard 99.999% N
2
(quality 50) or equivalent oxygen free gas
High range sensors: Standard 99.999% N
2
(quality 30) or equivalent oxygen free gas, air
Accessories
Accessories
Active spots, spare sensors, tool kit, portable calibration setup, etc.
Ask your local Hach Lange representative for more details on all available spare parts and
accessories.
Installation
In process sensor installation
devices
ORBISPHERE insertion/extraction valve for installation on Varinline® access units
ORBISPHERE 28 mm sensor weld-on socket
PG 13.5 stationary housing for installation on Varinline® access units
Note:
This device is only available for the M1100 sensor
28 mm stationary housing for installation on Varinline® access units
Ambient temperature
-5 to 50 °C (23 to 122 °F)
Humidity
0 to 95% non-condensing relative humidity
Power supply
Universal 85-264 VAC @ 50/60 Hz, 25 VA
10-36 VDC, 25 W
Weights
Wall (pipe) mount controller
3.8 kg
Panel mount controller
2.9 kg
M1100 12 mm (PG 13.5) sensor
0.6 kg
K1100 and M1100 28 mm sensor
0.74 kg
Calibration device
0.7 kg