Inputs/Outputs Menu
Analog outputs
There are three analog outputs available. These outputs are configurable in terms of function, content,
and behavior through the instrument menus. Analog outputs are used to output a current which is a
function (e.g. a linear characteristic) of a measurement: AOut = f (M). The analog outputs can be
typically connected to a PLC. Knowing the function (f), the PLC can compute the value of the
measurement.
Instrument
configuration
•
Select analog output range of current:
4-20 mA or 0-20 mA
•
The 4-20 mA range (recommended) allows for an extended event
indication mode that can be selected and configured (default = standard
mode)
For some events (sensor out, purge failure, etc.) the actual measurement is
not significant, but the PLC needs to know how the analog output behaves
in these cases. Two "Event indication modes" are available:
•
Standard mode (default)
•
Extended mode
Standard event
indication
Refer to
Table 1 Standard event indication
below.
Extended event
indication
The "Extended event indication" mode is only available when the 4-20 mA
output is selected. In this mode, the range between 0 mA and 4 mA is used
to indicate selected events. The events are defined using the channel
configuration option (see
Channel configuration on page 65
).
Table 1 Standard event indication
Analog output
Event output range
Event
0-20 mA
4-20 mA
Gas concentration
20 mA
20 mA
•
Channel out
•
Sensor out
•
Thermal cut-off
•
Interfering gas error
Temperature
20 mA
20 mA
•
Channel out
•
Sensor out