80
Communication Menu
Commands
The “Command Output Buffer” is formatted as follows:
The “Command Input Buffer” is located just after the measurement data and is formatted as
follows:
The following commands are available:
•
Change product
•
Activate sensor (valid for EC sensors only)
Change product command - output
Change product command - input
Name
Type
Size
Offset
Output command toggle (OCT)
Output byte
8 bits
0
Output command ID (OCI)
Output byte
8 bits
1
Output command data byte 1 (OCD1)
Output byte
8 bits
2
Output command data byte 2 (OCD2)
Output byte
8 bits
3
Output command data byte 3 (OCD3)
Output byte
8 bits
4
Output command data byte 4 (OCD4)
Output byte
8 bits
5
Name
Type
Size
Offset
Input command toggle (ICT)
Input byte
8 bits
74
Input command status (ICS)
Input byte
8 bits
75
Input command data byte 1 (ICD1)
Input byte
8 bits
76
Input command data byte 2 (ICD2)
Input byte
8 bits
77
Input command data byte 3 (ICD3)
Input byte
8 bits
78
Input command data byte 4 (ICD4)
Input byte
8 bits
79
Name
Value
Comment
OCT
1-2
OCI
1
OCD1
0
Channel number:
0 = Channel 1
OCD2
0-99
Product number
OCD3
0-1
Erase measurement files:
0 = Never erase the measurement files.
1 = Erase measurement file if necessary (e.g. gas unit changes)
OCD4
Not
used
Name
Value
Comment
ICT
1-2
ICS
0-3
0 = OK
1 = Unknown command ID
2 = Invalid parameter (e.g. invalid channel no or product number)
3 = Execution failure
ICD1
Not used
ICD2
Not used
ICD3
Not used
ICD4
Not
used