beautypg.com

Communication menu – Hach-Lange ORBISPHERE K1100_M1100 Operator Manual User Manual

Page 76

background image

74

Communication Menu

Gas sensor calibration report example

Calibration report nb 1
Calibration mode . . . . . . . . . . . .Manual high level adjustment
Date (yy.mm.dd - hh:mm) . . . . .05.02.17 - 18:40
Operator . . . . . . . . . . . . . . . . . .jp
Operator ID . . . . . . . . . . . . . . . .3
Cal. sample . . . . . . . . . . . . . . . .In line sample
Medium . . . . . . . . . . . . . . . . . . .Liquid
Gas unit. . . . . . . . . . . . . . . . . . .ppb
Liquid. . . . . . . . . . . . . . . . . . . . .Beer
Ref. value . . . . . . . . . . . . . . . . .1.500000
Cal. coefficient Phi0(Tcal) . . . . .25.974°
Cal. coefficient Tcal . . . . . . . . . .24.41°C
Cal. coefficient Ksv(25°C) . . . . .-0.1312 mbar-1
Cal. coefficient Tcal.Inst . . . . . .25.20°C
Standard deviation . . . . . . . . . .0.003 mbar
Fluor. amplitude. . . . . . . . . . . . .0.834 V
Interferences . . . . . . . . . . . . . . .Disable
Temperature . . . . . . . . . . . . . . .20.1 °C
Calibration duration . . . . . . . . . .3 mn

Barometric sensor calibration report example

Calibration report nb 1
Date (yy.mm.dd - hh:mm) . . . . .05.02.16 - 20:38
Operator . . . . . . . . . . . . . . . . . .jp
Operator ID . . . . . . . . . . . . . . . .3
Previous barometric pressure . .0.956 bar
New Barometric pressure . . . . .0.976 bar
Calibration offset . . . . . . . . . . . .3.73 [kPa]

Interfering gas calibration report example

Calibration report nb 1
Date (yy.mm.dd - hh:mm) . . . . .05.02.17 - 20:03
Operator . . . . . . . . . . . . . . . . . .jp
Operator ID . . . . . . . . . . . . . . . .3
Calibration Coefficient . . . . . . . .366.127661
Chlorinity/Salinity. . . . . . . . . . . .19.000000 [g/l]
Ratio with last calibration . . . . .99.990075 %
% of ideal current . . . . . . . . . . .111.624287 %
Stability . . . . . . . . . . . . . . . . . . .0.000000 %
Temperature . . . . . . . . . . . . . . .20.1 °C
Barometric pressure . . . . . . . . .1.020 bar
External pressure . . . . . . . . . . .0.977 bar
Sensor current. . . . . . . . . . . . . .7.990 µA

User action log file example

The "User action log file" below contains 3 user actions.

Nr

mm/dd hh:mm:ss User

ID

User Name

Action ID

Description

1 1/21

15:13:44

1007

Armstrong

139

Change Time/Date

0 1/21

15:13:27

1007

Armstrong

501

Cal. parameters

2 1/21

15:12:15

1007

Armstrong

132

Identification