Figure 27


c. Adjust the number until the DMM indicates the closest value to 15V.
d. Press the OK soft key to finish this step. The message “Step 1 done.” is
displayed on the screen.

8.2.2 Step 2

a. PB-507 will display:

Figure 28


c. Adjust the number until the DMM indicates the closest value to 1V.
d. Press the OK soft key to finish this step. The message “Step 2 done.” is
displayed on the screen.

8.2.3 Step 3

a. Connect the oscilloscope probe to the TTL and GND terminals of the Pulse
generator module. The TTL LED will turn on, the CMOS LED will turn off and PB-
507 will display:

Figure 29