Using the Configuration Software to Display and Set the Calibration values of the ST50/ST75
• From the Utilities Menu tap on the “Calibration Values” button to open the
calibration Parameters screen.
• Tap On the “Get All” button to read the current setup of the ST50/ST75.
• To Change the value of a Parameter, scroll up or down until you find the
parameter, tap on it to select it, then enter the new value in the value field.
• Repeat the previous step for as many parameters as need to be changed.
• When you are ready to down load the new values to the ST50/ST75 tap on the
“Set All” button.
Using the Configuration Software to Restore the ST50/ST75 Factory set Parameters
• From the Utilities Menu tap on the “Restore” button to
open the Factory Restore screen. By performing a restore,
the parameters that have changed in the field, will be lost,
and the ST50/ST75 Factory calibration setting will be
reloaded into the current active memory of the instrument.
• Tap on the “Yes – Restore” button to initiate the Factory
Restore.
• Tap on the “Return” button to return to the Utilities
Menu.
Using the Configuration Software to View the ST50/ST75 Process and System Faults
The ST50/ST75 has nine possible alarm conditions it can detect; 8 are process related and 1 is instrument related.
• Tap On the “Get All” button to read the current setup of the ST50/ST75. At
this time any fault conditions that exist will show with a check on the check
box.
• Process related Faults:
Below Min Flow
Above Max Flow
Below CMin Flow
Above CMax Flow
Below CMin Temp
Above CMax Temp
TC DeltaR Below DeltaR Min
TC DeltaR Above DeltaR Max
•
Instrument related Fault: Board Temp Out of Limit.