beautypg.com

Burkert Type 3005 User Manual

Page 17

background image

17

Installation

procedure:

Fig. 10: Standard installation

Ensure that the ball valve / the flap valve is in its closed position.

Carefully connect the rotary actuator to the shaft of the ball valve
/ flap valve.

When connecting the rotary actuator, ensure that it is not
twisted and that the fastening threads of the rotary actuator
are covered by the fastening bores of the ball valve / flap
valve.

Screw the fastening screws into the fastening threads of the rotary
actuator and tighten them firmly (max. 3 Nm).

8.2.2. set mechanical end position limit

(actuators 100 - 1000 nm)

The mechanical end position limits have been preset at
the factory and glued on with Loctite. However, they can
be adjusted by turning the screws M8 Pos. 17. Then the
nuts must be glued on again.

Loosen the M8 nuts on the mechanical limit stops 17 (see “Fig. 1”
and “Fig. 3”) and adjust the mechanical end position limits.

Glue nuts on again with Loctite (e.g. Loctite 577).

8.2.3. adjusting limit switch contacts

The two upper limit switch contacts have been set to
0 - 90° at the factory.

removing position indicator and hood

Fig. 11: Removing

glass hood

Fig. 12: Removing

position
indicator

Fig. 13: Removing

hood

procedure:

Remove glass hood of the position indicator 1 including the sealing
ring by loosening the four fastening screws and remove the glass
hood (see “Fig. 11”).

Remove position indicator (see “Fig. 12”).

Remove hood 2 by loosening the four fastening screws 3 (see
“Fig. 13”).

english

Type 3005