Details of the process level, Details of the parameters menu – Burkert Type 8025 User Manual
Page 49
49
operatingandcommissioning
Type 8025 / 8035
9.5.
Details of the process level
This level is active by default when the device is energized.
12.6 L/MiN.
16.45 MA
87654 L
231 L.
0......9
> 2 s
Resetting the daily totalizer.
Value of the measured flow rate, displayed in the unit chosen in the "UNIT" parameter of the
Parameters menu.
Value of the current output, proportional to the measured flow rate.
Value of the main totalizer, volume of fluid counted by the device since the last reset.
Value of the daily totalizer (identified by a dot after the volume units), volume of fluid counted by
the device since the last reset.
0......9
+
Fig. 53: Diagram of the Process level
9.6.
Details of the parameters menu
To access the Parameters menu, simultaneously press keys
ENTER
for at least 5
s.
This menu comprises the following configurable parameters:
LANGUAGE
Process level
UNit
K-FACtOR
CURRENt
FiLtER
tOtAL
PULSE
RELAy
END
12.6 L/MiN.
0......9
Choosing the display language
Choosing the flow rate unit, the number of decimals and the unit the totalizers are displayed in.
Entering the K factor of the fitting used or have it defined through a teach-in procedure.
Choose the flow rate range related to the 4-20 mA output current.
Choosing the filter level of the measured flow rate, on the displayed flow rate and the current
output. Ten dampening levels are available.
Resetting both totalizers.
Parameterizing the pulse output.
Configuring the relays. This function only appears on devices with relays.
Saving the changes made within the Parameters menu and returning
to the Process level.
Fig. 54: Diagram of the Parameters menu
English