Details of the process level, Details of the parameters menu – Burkert Type 8025 User Manual
Page 49

49
operatingandcommissioning
Type 8025 / 8035
9.5.
Details of the process level
This level is active by default when the device is energized.
12.6 L/MiN.
16.45 MA
87654 L
231 L.
0......9
> 2 s
Resetting the daily totalizer.
Value	of	the	measured	flow	rate,	displayed	in	the	unit	chosen	in	the	"UNIT"	parameter	of	the	
Parameters	menu.
Value of the current output, proportional to the measured flow rate.
Value of the main totalizer, volume of fluid counted by the device since the last reset.
Value	of	the	daily	totalizer	(identified	by	a	dot	after	the	volume	units),	volume	of	fluid	counted	by	
the	device	since	the	last	reset.
0......9
+
Fig. 53: Diagram of the Process level
9.6.
Details of the parameters menu
To access the Parameters menu, simultaneously press keys
ENTER
for at least 5
s.
This menu comprises the following configurable parameters:
LANGUAGE
Process level
UNit
K-FACtOR
CURRENt
FiLtER
tOtAL
PULSE
RELAy
END
12.6 L/MiN.
0......9
Choosing the display language
Choosing the flow rate unit, the number of decimals and the unit the totalizers are displayed in.
Entering the K factor of the fitting used or have it defined through a teach-in procedure.
Choose the flow rate range related to the 4-20 mA output current.
Choosing	the	filter	level	of	the	measured	flow	rate,	on	the	displayed	flow	rate	and	the	current	
output.	Ten	dampening	levels	are	available.
Resetting both totalizers.
Parameterizing the pulse output.
Configuring the relays. This function only appears on devices with relays.
Saving	the	changes	made	within	the	Parameters	menu	and	returning	
to	the	Process	level.
Fig. 54: Diagram of the Parameters menu
English