English, T 125 ma – Burkert Type 8025 User Manual
Page 35
35
Wiring
Type 8025 / 8035
8.4.5. Wiring the 8025 compact version and the 8035, 115/230 V Ac, without
relays, with cable glands
→
Before wiring the device, configure the selector on the electronic board (see chap. "8.3").
The red wire is wired on the NC terminal to make the wiring of the
4-20 mA output easier (see "Fig. 34"). The jumper wire 4 between L+ and
NC energizes the NC terminal.
Supply
12..36Vdc
FLOW SENSOR
COIL
NPN
L+
L-
PE
P-
P+
NC
L+
L-
PE
P-
P+
Iout
PULSE
OUTPUT
Without
With
Relays
230V
T 125 mA
N
PE
L
{
3
5
6
7
4
A
2
1
B
PE
Switch A : see chap. "8.3.1"
Switch B : see chap. "8.3.3"
terminal block 1
NC: not connected (terminal for the
wiring of the 4-20 mA output)
L+ (red wire, factory wired)
PE: wiring of the PE between the main
board and the protection board
L- (black wire, factory wired)
P-: negative pulse output
P+: positive pulse output
terminal block 2 pe
Shield wiring (green/yellow wire,
factory wired)
connector 3: connection of the flow
sensor
Jumper wire 4: jumper wire to be
removed if the 4-20 mA output is
connected
time-delay fuse 5: fuse to protect the
115 V AC or 230 V AC power supply
earth terminal 6 of the housing:
internally connected to the earth plug
terminal block 7: wiring of the
115/230 V AC power supply
Fig. 33: Terminal assignment of a 115/230V AC compact version without relays, with cable glands
Supply
12..36Vdc
FLOW SENSOR
COIL
NPN
L+
L-
PE
P-
P+
NC
L+
L-
PE
P-
P+
Iout
PULSE
OUTPUT
Without
With
Relay
s
230V
T 125 mA
N PE
L
{
+ -
PE
4-20mA input at
external instrument
115/230 V AC power supply
If the current output is wired, remove the jumper wire between the terminals Iout and L+.
Fig. 34: Wiring of the current output of a compact version, 115/230 V AC, without relays, with cable glands
English