beautypg.com

Model 0254 – Brooks Instrument 0254 User Manual

Page 94

background image

C-16

Model 0254

Section C Serial Communications Protocol

Installation and Operation Manual

X-SE-0254-eng

Part Number: 541B129AAG

September, 2010

C-4-4 Global Settings Services

C-4-4-1 Global Setting Values

This command enables a terminal operator or host to acquire the present
state of the Global programmed values. The Global Settings port number
is one greater than the maximum number of available ports, which is 9 for
Model 0254.

HOST SEND

AZ[.09]V

Non-Network

AZ[.9] V

AZ[yyyyy.09]V

Network

AZ[yyyyy.9]V

RESPONSE

PROGRAM VALUES - Channel Global

<39> Audio Beep

On

<32> Zero Supress

On

<33> Pwr SP Clear

Off

<43> Record Count

000000 (view only)

log option

<25> Sample Rate

535 sec

log option

<22> Date-Time

00Jan00 00:00:00

log option

<17> Network Addr

00000

C-4-4-2 Set Global Settings Values

Each of the system port programmed operating values can be individually
queried or changed in accordance with the procedures in Section 3 using
the value index.

C-4-5 Communication Message Basics

Messages between host and instrument are either polled (solicited) or un-
polled (un-solicited), where the host is normally the polling (soliciting) party.