beautypg.com

Model 0254, Appendix b blending examples – Brooks Instrument 0254 User Manual

Page 77

background image

B-1

Model 0254

Appendix B Blending Examples

Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010

B-1 Blending Examples

Example 4:

Master channel flow is 100%
Slave channel flow has to be 50% (of slave channel full
scale)

Mass flow controller selections are :

Master channel unit: 6 l/min = 100% Full Scale
Slave channel unit: 4 l/min = 100% Full Scale

Blending ratio is = 50%

NOTE: When the blend ratio is % of slave channel, use
100% as the full scale setting for master and slave.

Example 5:

Master channel flow is 80%
Slave channel flow has to be 20 kg/hr

Mass flow controller selections are :

Master channel unit: 100 kg/hr =100% Full Scale
Slave channel unit: 50kg/hr = Full Scale Value

Determination of entered setpoint:

20 kg/h
Blending ratio is = = 0.25 = 25%

80%

kg/h

Setpoint

slave

= 0.25 x 100% = 25 kg/h

%

Example 6:

Master channel pressure is 10 bar
Slave channel flow has to be 50% (of master channel full
scale)

Mass flow controller selections are :

Master channel unit: 15 bar = Full Scale Value
Slave channel unit: 20 l/min = Full Scale Value

Blending ratio is = 50%

Example 1:

Master channel flow is 80 l/min
Slave channel flow has to be 0.8 l/min

Mass flow controller selecions are :

Master channel unit: 100 l/min = Full Scale Value
Slave channel unit: 10 l/min = Full Scale Value

0.8 l/min
Blending ratio is = = 0.01 = 1%
80 l/min

Example 2:

Master channel flow is 240 l/min
Slave channel flow has to be 40 l/min

Mass flow controller selections are :

Master channel unit: 300 l/min = Full Scale Value
Slave channel unit: 50 l/min = Full Scale Value

40 l/min
Blending ratio is = = 0.167 = 16.7%
240 l/min

Example 3:

Master channel pressure is 1000 mbar
Slave channel flow has to be 50 l/min

Mass flow and pressure controller selections are :

Master channel unit: 2000 mbar = Full Scale Value
Slave channel unit: 100 l/min = Full Scale Value

50 l/min
Blending ratio is = = 0.05 = 5%

1000 mbar