background image

C-12

Model 0254

Section C Serial Communications Protocol

Installation and Operation Manual

X-SE-0254-eng

Part Number: 541B129AAG

September, 2010

C-4-2-2 Send Channel Input Port Programmed Values

This command enables a terminal operator or host to acquire the present
state of input port programmed values. Valid input port numbers (xx) are
[1, 3, 5, 7].

HOST SEND

AZ[.xx]V

Non-network

AZ[yyyyy.xx]V

Network

RESPONSE

PROGRAM VALUES - Channel 1 - Port 01

<04> Measure Units ml

<10> Time Base min

<03> Decimal Point x.xx

<27> Gas Factor 1.000

<28> Log Type Off

<00> PV Signal Type 0-20mA

<09> PV Full Scale 20.00 ml/m

Each of the input port programmed operating values can be individually
queried or changed in accordance with the procedures in Section C-5-1