beautypg.com

B&K Precision 9130 - Manual User Manual

Page 37

background image

37

STATus:OPERation:INSTrument:CONDition?

This command queries parameters from operation condition sub register. After this command is executed,

the operation condition sub register is reset.

Note: this command is only valid for current channel.

Command syntax: STATus:OPERation:INSTrument:CONDition?

Return parameter:

Example: STAT:OPER:INST:COND?

INSTrument[SELect]

This command can be used to select the channel.

Command syntax: INSTrument[:SELect]

Parameter: FIRst|SECOnd|THIrd

Reset value: FIRst

Example: INST SECO

INSTrument[:SELect]?

This command returns the number of the selected channel

Command syntax: INSTrument[:SELect]?

Return parameter: FIRst|SECOnd|THIrd

Example: INST?

INSTrument:NSELect

This command is similar to the command “INSTrument[SELect]”, the only difference is that this command

uses numbers to denote the channel.

Command syntax: INSTrument:NSELect

Parameter: 1~3

Reset value: 1

Example: INST:NSEL?

OUTPut[:STATe]

This command sets the output state for the current channel to ON or OFF

Command syntax: OUTPut[:STATe]

Parameter: 1(ON)|0(OFF)

Example: OUTP 1

OUTPut[:STATe]?

This command queries the output state of the current channel is ON or OFF

Command syntax: OUTPut[:STATe]?

Return parameter: 0|1

Example: OUTP?

OUTPut:TIMer:DATA

This command sets the time of the output-timer for the current channel.

Command syntax: OUTPut:TIMer:DATA