B&K Precision 9130 - Manual User Manual
Page 10

10
14) Verify channels 2 & 3 according to steps 1) – 8)
5.If the Power Supply Does Not Turn On
Use the following steps to help resolve problems you might encounter when turning on the instrument.
1). Verify that there is AC power applied to the power supply.
Verify that the power cord is firmly plugged into the power receptacle on the rear panel of the power supply.
Make sure the power outlet you are using is working properly and verify that the power supply is turned on.
2). Verify the power-line voltage setting.
Make sure the voltage selector switch is set according to the present line voltage (110VAC or 220VAC).
Change the voltage setting if it’s not correct.
3). Verify that the correct power-line fuse is installed.
Model
Fuse Description
9130
Fuse 3.15A T250V(220V AC)
Fuse 6.30A T250V(110V AC)
If the fuse is blown replace it according to the table above.
6.To Adjust the Carrying Handle
To adjust the position, grasp the handle on each side and pull outwards, then rotate the handle to the desired
position.
Bench-top viewing positions Carrying position
7. To Rack Mount the Instrument
You can mount the power supply in a standard 19-inch rack cabinet using the IT-E151 rack mount kit.
Note: Remove the carrying handle and the two plastic ears before rack-mounting the instrument. To remove
the handle, grasp the handle on the side, pull outwards and rotate it to a special position where the